Computational fluid dynamics simulation of Reynolds stress frequencies in the FDA nozzle


Avci M.

Niğde Ömer Halisdemir Üniversitesi Mühendislik Bilimleri Dergisi, cilt.13, sa.3, ss.969-974, 2024 (Hakemli Dergi) identifier

Özet

It is known that examining turbulence effects on medical devices has an important effect in design and optimization of blood-contacting devices. CFD has been commonly used on prosthetic heart valves, stents, and Ventricular Assist Devices (VADs) in both the design process and also on hemodynamics of the flow characteristics. In this study, flows in the FDA nozzle were modeled to examine Reynolds stresses in the whole domain. The flow behavior was determined by applying the Reynolds-Averaged Navier-Stokes model of turbulence (k-ω SST) to simulate five distinctive experimental cases in the nozzle taken from the literature. The Reynolds stress frequencies are determined for the five different experimental conditions. Results showed that the highest velocity case (corresponding throat Reynolds number of 6500) has much higher Reynolds stresses with a high number of frequencies. However, the lowest velocity case has very small Reynolds numbers in a very high frequency. When different parts of the nozzle were examined, the Reynolds stress values showed more fluctuations for the higher velocities and more regular profiles for the lower velocity cases.